Fluxmeter Measurement System |
Features
|
|
Description
The Model 2130 Fluxmeter is a fully featured, high end
Fluxmeter based measurement system. It is a compact instrument with high accuracy,
automatic drift correction, microprocessor control and twelve measurement ranges to allow
the operator to configure the meter for maximum resolution and accuracy. The Model 2130 is
capable of measuring (DC), alternating (AC), or pulsed magnetic fields. Selecting
primary measurement functions is quick and easy with dedicated front panel function keys.
An easy to use menu allows access to user defined parameters such as upper and lower
magnetic limits and relative magnetic measurements. Remote meter configuration is
completely accessible using the built-in RS-232 Interface or optional IEEE-488 Interface.
The Model 2130 is an electronic integrating Fluxmeter with high sensitivity, very low
drift and automatic drift correction. It provides measurements in five (5) different units
over twelve (12) measurement ranges. The ability to select measurements in either Maxwell,
Maxwell-turns, Gauss, Tesla and Bdi Gauss make this unit ideal for both production and
laboratory/research testing. The most frequently used functions are accessed using front
panel pushbuttons while more complex functions are selected by way of a user friendly menu
system and function select entry. For example, Helmholtz Coil constants and magnet volume
can be manually entered from the front panel and stored in memory to provide readings
directly in Bdi Gauss. Probe/Coil parameters, such as the number of turns and area (NA),
resistance and Helmholtz constant, can also be stored in look-up tables for easy reference
and future setups. This feature saves time and reduces the chance of errors in converting
readings manually.
Magnetic Instrumentation, Inc. offers a wide selection of both standard and custom
Helmholtz Coil Systems and Search Coils to measure many different magnet sizes and
configurations.
Specifications
Accuracy Specifications:
| Function | Range | Typical (% of range) |
Maximum (% of range) |
Conditions |
| Normal (DC) | 105- 108 | 0.050 | 0.100 | - |
| AC (RMS) | 105 - 108 | 0.500 | 1.000 | 20 Hz - 30 Hz |
| AC (RMS) | 105 - 108 | 0.300 | 0.700 | 30 Hz - 1 k Hz |
| Max Hold (Peak) | 105 - 108 | 0.300 | 0.800 | ³ 1 ms pulse width * |
| Analog Output | 105 - 108 | 0.150 | 0.400 | Specification is for the deviation of the analog output compared to the Model 2130 display value. |
* Certified at a 1ms pulse. However, if the input voltage range is not exceeded impulse fields as fast as 100 microseconds can be measured.
| Average Time Constant:
(display rate) This interval may be set from 100 ms to 2500 ms, in 100 ms intervals. (0.4 to 10 times per second) Display Resolution: 4 3/4 Digits (1 part out of ± 32,000) Operating Temperature Range: 0ºC to +50ºC Storage Temperature Range: -20ºC to +70ºC |
Maximum Resolution: 10
Maxwell-turns on 105 range Humidity: 0 to 68% Analog Output: 1 Volt ± 0.25% of full scale Power Requirements: 110 VAC, 220 VAC at 10W, or ±9 VDC at 5W Operation Dimensions: Length 12.00 in (305 mm), Width 6.70 in (170 mm), Height 2.25 in (57 mm), Approximate Weight: 6 lb (2.7 kg) |
Specifications subject to change without notice.